Recall Ledger

HemoslL AcuStar HIT-lgG recall (Instrumentation Laboratory)

Medical deviceClass IOngoing

Summary: Instrumentation Laboratory is recalling HemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009800127. in vitro diagnostic. Reason: Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples.

What is being recalled

HemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009800127. in vitro diagnostic

Why

Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples.

How serious is it

Class I is the most serious level: FDA considers there is a reasonable probability that using or being exposed to this product will cause serious health problems or death.

Where it was sold

Worldwide - US Nationwide distribution in the states of AR, AZ, CA, CT, FL, GA, MA, MD, NJ, NY, PA, RI, TN, WA, WV and the countries of Austria, Belgium, Switzerland, Costa Rica, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, India, Italy, Jordan, Japan, South Korea, Luxembourg, Latvia, Netherlands, Saudi Arabia, Sweden, Singapore. (states named: Arkansas, Arizona, California, Connecticut, Florida, Georgia, Massachusetts, Maryland, New Jersey, New York, Pennsylvania, Rhode Island, Tennessee, Washington, West Virginia)

How to tell if yours is affected

Part Number: 0009800127. UDI-DI: 08426951126545. Lot Number (Expiration Date): B40005 (01/16/2027), B38062 (11/01/2026).

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-0001-2027
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass I
StatusOngoing
Recall startedSeptember 3, 2026
ReportedOctober 21, 2026
CompanyInstrumentation Laboratory
Company locationBedford, MA, United States
Initiated byVoluntary: Firm initiated
Quantity434 units