Brand Name: MiniMed Flex Product Name: MiniMed Flex Insulin Pump Model/Catalog... recall (Medtronic MiniMed, Inc.)
Medical deviceClass IIOngoing
Summary: Medtronic MiniMed, Inc. is recalling Brand Name: MiniMed Flex Product Name: MiniMed Flex Insulin Pump Model/Catalog Numbers: MMT-8062; MMT-8072 Product Description: The MiniMed Flex pump is an alternate controller enabled (ACE) pump intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diab. Reason: Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed verification test results in one or more of 9 functional tests. The unverified/failed pumps released to users potentially have one or more of the following issues: inoperable pump at setup, lack of protection for moisture exposure, slow battery charging, disconnection issues with other system components (i.e., continuous glucose monitoring sensor or MiniMed mobile app), reduced ability to detect occlusion events, compromised audio for pump alarms/alerts, and/or inaccurate insulin delivery.
What is being recalled
Brand Name: MiniMed Flex Product Name: MiniMed Flex Insulin Pump Model/Catalog Numbers: MMT-8062; MMT-8072 Product Description: The MiniMed Flex pump is an alternate controller enabled (ACE) pump intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. It can reliably and securely communicate with compatible digitally connected devices, including an integrated continuous glucose monitor (iCGM), interoperable Medtronic continuous glucose monitor (CGM), and interoperable automated glycemic controller (iAGC). The pump is intended to be used in conjunction with compatible, digitally connected medical devices for the purpose of drug delivery. The primary user interface and primary display for the MiniMed Flex pump is the MiniMed app, also the user interface for the compatible MiniMed iAGCs, which runs on a compatible iOS or Android device.
Why
Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed verification test results in one or more of 9 functional tests. The unverified/failed pumps released to users potentially have one or more of the following issues: inoperable pump at setup, lack of protection for moisture exposure, slow battery charging, disconnection issues with other system components (i.e., continuous glucose monitoring sensor or MiniMed mobile app), reduced ability to detect occlusion events, compromised audio for pump alarms/alerts, and/or inaccurate insulin delivery.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
U.S. Nationwide distribution in the states of AR, AZ, CO, FL, KS, KY, LA, MI, MN, NJ, NY, OH, PA, SD, TN, TX, and WV. (states named: Arkansas, Arizona, Colorado, Florida, Kansas, Kentucky, Louisiana, Michigan, Minnesota, New Jersey, New York, Ohio, Pennsylvania, South Dakota, Tennessee, Texas, West Virginia)
How to tell if yours is affected
MMT-8072 Lot Code: Serial numbers: FB00015839, FB00016511, FB00014455, FB00011553, FB00015700, FB00011357, FB00019307, FB00014766, FB00014822, FB00015828, FB00018580, FB00010465, FB00014472, FB00015793, FB00015794, FB00015796, FB00015799, FB00015801, FB00015977, FB00017271. UDI-DI/GTIN: 00763000848156 Note: Initially 25 pumps were identified as put into service, however 2 were shipped to Marketing Team Members and recovered. 3 Pumps were returned through complaint process. Final total of pumps available in the field is 20. MMT-8062 Lot Code: Serial numbers: FB00011770, FB00015162, FB00016429, FB00014760, FB00015797 UDI-DI/GTIN: 00763000848149 (Serial Numbers: FB00017271, FB00011770, FB00015162, FB00016429, FB00014760 were returned via complaint process or shipped to Marketing Team Members).
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-0088-2027 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | August 20, 2026 |
| Reported | October 14, 2026 |
| Company | Medtronic MiniMed, Inc. |
| Company location | Northridge, CA, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 25 |
More recalls from Medtronic MiniMed, Inc. · all medical device recalls · recalls from August 2026