Recall Ledger

TufTex¿ Over-the-Wire Embolectomy Catheter recall (LeMaitre Vascular, Inc.)

Medical deviceClass IIOngoing

Summary: LeMaitre Vascular, Inc. is recalling TufTex¿ Over-the-Wire Embolectomy Catheter; Catalog Number: E1651-88;. Reason: Devices failed in process testing related to the amount of pull force it can sustain prior to the balloon components failing.

What is being recalled

TufTex¿ Over-the-Wire Embolectomy Catheter; Catalog Number: E1651-88;

Why

Devices failed in process testing related to the amount of pull force it can sustain prior to the balloon components failing.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

US Nationwide distribution in the states of CA, FL, GA, KY, LA, NJ, OH, PA, SC. (states named: California, Florida, Georgia, Kentucky, Louisiana, New Jersey, Ohio, Pennsylvania, South Carolina)

How to tell if yours is affected

Catalog Number: E1651-88; UDI-DI: 00840663100866; Lot Number: XOT00002181;

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-0117-2027
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedAugust 28, 2026
ReportedOctober 14, 2026
CompanyLeMaitre Vascular, Inc.
Company locationBurlington, MA, United States
Initiated byVoluntary: Firm initiated
Quantity54

More recalls from LeMaitre Vascular, Inc. · all medical device recalls · recalls from August 2026