MAS Omni IMMUNE PRO Liquid Assayed Integrated Immunoassay Control recall (Microgenics Corporation)
Medical deviceClass IIOngoing
Summary: Microgenics Corporation is recalling MAS Omni IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, REF: OPRO-101; MAS Omni IMMUNETM PRO Liquid Assayed Integrated Immunoassay Control, Sample Pack, REF: OPRO-SP. Reason: Vial-to-vial variability in MAS¿ Omni"IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, Lot OPRO29061A results in low Parathyroid Hormone (PTH) recovery, which may lead to a delay in reporting patient PTH results.
What is being recalled
MAS Omni IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, REF: OPRO-101; MAS Omni IMMUNETM PRO Liquid Assayed Integrated Immunoassay Control, Sample Pack, REF: OPRO-SP
Why
Vial-to-vial variability in MAS¿ Omni"IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, Lot OPRO29061A results in low Parathyroid Hormone (PTH) recovery, which may lead to a delay in reporting patient PTH results.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
Worldwide - US Nationwide distribution in the states of UT, MI, OK, NY, TX, CA, LA, SC, OH, MO, OR and the countries of KOR, BGD, THA, ZAF, DEU, VNM, BRA, NZL. (states named: California, Louisiana, Michigan, Missouri, New York, Ohio, Oklahoma, Oregon, South Carolina, Texas, Utah)
How to tell if yours is affected
REF/UDI-DI/Lot(Expiration): OPRO-101/00884883005327/OPRO29061A(2029-06-30); OPRO-SP/00884883005358/OPRO2906SA(2029-06-30)
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-0159-2027 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | August 17, 2026 |
| Reported | October 21, 2026 |
| Company | Microgenics Corporation |
| Company location | Fremont, CA, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 1366 |