Williams Cystoscopic Injection Needles recall (Cook Incorporated)
Medical deviceClass IIOngoing
Summary: Cook Incorporated is recalling Williams Cystoscopic Injection Needles, used for cystoscopic injection into the urethra, bladder neck, and bladder wall, Reference Part Number 090001-S26, Order Number (GPN) G16112. Reason: The female Luer lock adapter component in the affected device lots may not have been correctly seated due to misalignment and inadequate torquing during manufacturing. Potential events that may occur as a result of this issue include increased procedural time and inadequate treatment.
What is being recalled
Williams Cystoscopic Injection Needles, used for cystoscopic injection into the urethra, bladder neck, and bladder wall, Reference Part Number 090001-S26, Order Number (GPN) G16112
Why
The female Luer lock adapter component in the affected device lots may not have been correctly seated due to misalignment and inadequate torquing during manufacturing. Potential events that may occur as a result of this issue include increased procedural time and inadequate treatment.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
US States: FL, IL, NM, TX, WA. Canada, France, Germany, Italy, Spain, Switzerland. (states named: Florida, Illinois, New Mexico, Texas, Washington)
How to tell if yours is affected
UDI-DI: 00827002161122; Lot Numbers: 16613606, 16951548, 16951548X
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-0163-2027 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | September 16, 2026 |
| Reported | October 21, 2026 |
| Company | Cook Incorporated |
| Company location | Bloomington, IN, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 97 units |