Recall Ledger

Brand Name: VACUETTE Product Name: LH Lithium Heparin Separator Model/Catalog... recall (Greiner Bio-One North America, Inc.)

Medical deviceClass IIOngoing

Summary: Greiner Bio-One North America, Inc. is recalling Brand Name: VACUETTE Product Name: LH Lithium Heparin Separator Model/Catalog Number: 454008P Product Description: Greiner VACUETTE Blood Collection Tube with Lithium Heparin and Gel Separator (13 x 75 mm, 3.5ml). Reason: Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased calcium (Ca++) results which have the potential to cause hyperkalemia.

What is being recalled

Brand Name: VACUETTE Product Name: LH Lithium Heparin Separator Model/Catalog Number: 454008P Product Description: Greiner VACUETTE Blood Collection Tube with Lithium Heparin and Gel Separator (13 x 75 mm, 3.5ml)

Why

Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased calcium (Ca++) results which have the potential to cause hyperkalemia.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

U.S. Nationwide distribution in the states of CO, FL, IL, MO, NE, NY, OH, PA, and VA. (states named: Colorado, Florida, Illinois, Missouri, Nebraska, New York, Ohio, Pennsylvania, Virginia)

How to tell if yours is affected

Lot Code: Item Number: 454008P Lot Number: B26043FM UDI: Rack: 29120017579032, Case: 39120017579039 Expiration Date: 2027-07-12

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-0200-2027
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedJuly 29, 2026
ReportedOctober 28, 2026
CompanyGreiner Bio-One North America, Inc.
Company locationMonroe, NC, United States
Initiated byVoluntary: Firm initiated
Quantity860450

More recalls from Greiner Bio-One North America, Inc. · all medical device recalls · recalls from July 2026