Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common... recall (AGFA Healthcare Corp.)

Medical deviceClass IIOngoing

Summary: AGFA Healthcare Corp. is recalling Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name: DR 800. Reason: It is possible to exceed the maximum allowed Air Kerma rate (AKR) of 88mGy/min for pulsed fluoro exams. This can happen in case the framerate is increased on the NX workstation when ABS is not enabled.

What is being recalled

Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name: DR 800

Why

It is possible to exceed the maximum allowed Air Kerma rate (AKR) of 88mGy/min for pulsed fluoro exams. This can happen in case the framerate is increased on the NX workstation when ABS is not enabled.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

U.S. Nationwide distribution.

How to tell if yours is affected

N/A

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-0923-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedNovember 18, 2025
ReportedDecember 31, 2025
CompanyAGFA Healthcare Corp.
Company locationGreenville, SC, United States
Initiated byFDA Mandated
Quantity35