Recall Ledger

BD Pyxis MedStation ES recall (CareFusion 303, Inc.)

Medical deviceClass IIOngoing

Summary: CareFusion 303, Inc. is recalling BD Pyxis MedStation ES, REF: 323, BD Pyxis MedStation ES 7 Drawer Auxiliary, REF: 324, BD Pyxis MedStation ES Tower, REF: 352, BD Pyxis Pro 7-Drawer Auxiliary, REF: 1149-00, BD Pyxis Pro 7- Drawer Auxiliary with Secure Tower, REF: 1152-00, BD Pyxis Pro MedStation Main, REF: 1155-00, BD. Reason: Automated Dispensing Cabinets (ADC) received a Half Height CUBIE drawer firmware update, which caused a Cubie Insert event which prompts software to generate a duplicate address, which causes an error and the drawer to fail, which may lead to inability or delay in accessing stored items, delay in the replenishment of ADCs or patient specific medications stored outside of ADCs.

What is being recalled

BD Pyxis MedStation ES, REF: 323, BD Pyxis MedStation ES 7 Drawer Auxiliary, REF: 324, BD Pyxis MedStation ES Tower, REF: 352, BD Pyxis Pro 7-Drawer Auxiliary, REF: 1149-00, BD Pyxis Pro 7- Drawer Auxiliary with Secure Tower, REF: 1152-00, BD Pyxis Pro MedStation Main, REF: 1155-00, BD Pyxis CII Safe ES, REF: 1116-00

Why

Automated Dispensing Cabinets (ADC) received a Half Height CUBIE drawer firmware update, which caused a Cubie Insert event which prompts software to generate a duplicate address, which causes an error and the drawer to fail, which may lead to inability or delay in accessing stored items, delay in the replenishment of ADCs or patient specific medications stored outside of ADCs.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide - US Nationwide distribution including in the states of CO, IN, OH, TX, KS, FL, GA, NC, CA, HI, IL, NY, MI, VA, AL, LA, NE, KY, OK, AR, TN, WY, NJ, WI, SD, AZ, NV, MO, MA, ID, WV, DE, PA, WA, MT, CT, IA, MD, PR, MN, NM, OR, VT, MS, ME, AK, UT, DC, GU, SC, NH, RI, ND, AS and the countries of CH, CA, KW, QA, NL, GB, UY, AU, NZ, BH, TH, SG, CN, FR, LB, BS, HK, TR, PL, AR, BR, CL, CO, MX, ES, BM, SA, IL. (states named: Alaska, Alabama, Arkansas, Arizona, California, Colorado, Connecticut, District of Columbia, Delaware, Florida, Georgia, Hawaii, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Maine, Michigan, Minnesota, Missouri, Mississippi, Montana, North Carolina, North Dakota, Nebraska, New Hampshire, New Jersey, New Mexico, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Vermont, Washington, Wisconsin, West Virginia, Wyoming)

How to tell if yours is affected

Firmware Version: 1.48. REF/UDI-DI: 323/10885403512667, 324/10885403512681, 352/10885403512674, 1149-00/10885403517822, 1152-00/10885403517839, 1155-00/10885403521300, 1116-00/10885403512605

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-0962-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedNovember 20, 2025
ReportedDecember 31, 2025
CompanyCareFusion 303, Inc.
Company locationSan Diego, CA, United States
Initiated byVoluntary: Firm initiated
Quantity53669

More recalls from CareFusion 303, Inc. · all medical device recalls · recalls from November 2025

Official FDA record