BVI Leos Laser and Endoscopy System. Model Number: OME6000U. recall (Beaver-Visitec International, Inc.)

Medical deviceClass IIOngoing

Summary: Beaver-Visitec International, Inc. is recalling BVI Leos Laser and Endoscopy System. Model Number: OME6000U.. Reason: Manufacturing deviation in which the ferrites specified for the USB cable assembly linking the Nurse Monitor touchscreen to the System Controller were not installed as intended. The omission of the ferrites on the USB communication cable results in non-conformance to the design specifications, which require two ferrites (one at each end of the cable).

What is being recalled

BVI Leos Laser and Endoscopy System. Model Number: OME6000U.

Why

Manufacturing deviation in which the ferrites specified for the USB cable assembly linking the Nurse Monitor touchscreen to the System Controller were not installed as intended. The omission of the ferrites on the USB communication cable results in non-conformance to the design specifications, which require two ferrites (one at each end of the cable).

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

US Nationwide distribution in the states of FL, MI, OK, TX, WI. (states named: Florida, Michigan, Oklahoma, Texas, Wisconsin)

How to tell if yours is affected

Model Number: OME6000U. UDI: (01)00886158037576(11)250513(21). Serial Numbers: 77UBV500003, 77UBV500004, 77UBV500005, 77UBV500006, 77UBV500010.

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-1008-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedDecember 5, 2025
ReportedJanuary 7, 2026
CompanyBeaver-Visitec International, Inc.
Company locationWaltham, MA, United States
Initiated byVoluntary: Firm initiated
Quantity5 units