Recall Ledger

MOSAIQ Oncology Information System with Particle Therapy License recall (Elekta, Inc.)

Medical deviceClass IIOngoing

Summary: Elekta, Inc. is recalling MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO). Reason: Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial treatment,Partial treatment occurs followed by treatment field interruptions with manually recorded partial treatment,Then treatment unit doesn't check meterset value so partial treatment delivered twice.

What is being recalled

MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)

Why

Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial treatment,Partial treatment occurs followed by treatment field interruptions with manually recorded partial treatment,Then treatment unit doesn't check meterset value so partial treatment delivered twice

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

US Nationwide distribution in the state of TX. (states named: Texas)

How to tell if yours is affected

MOSAIQ v 3.10.200, and 3.1.1.0 up to and including v 3.2.3.1 with Particle Base. Software Versions: 3.10.200, 3.1.1.0, 3.1.2.0, 3.1.3.0, 3.1.3.1, 3.1.3.2, 3.1.3.3, 3.1.3.4, 3.2.0.0, 3.2.1.0, 3.2.1.1, 3.2.1.2, 3.2.1.3, 3.2.1.4, 3.2.2.0, 3.2.2.1, 3.2.3.0, 3.2.3.1 UDI-DI: 07340201500026, 07340201500071

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-1078-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedDecember 10, 2025
ReportedJanuary 21, 2026
CompanyElekta, Inc.
Company locationAtlanta, GA, United States
Initiated byVoluntary: Firm initiated
Quantity1

More recalls from Elekta, Inc. · all medical device recalls · recalls from December 2025

Official FDA record