Asserachrom HPIA. Model/Catalog Number: 00615. The Asserachrom HPIA kit is... recall (Diagnostica Stago, Inc.)
Medical deviceClass IIOngoing
Summary: Diagnostica Stago, Inc. is recalling Asserachrom HPIA. Model/Catalog Number: 00615. The Asserachrom HPIA kit is intended for the qualitative detection in plasma or serum of the anti-heparin-platelet factor 4 (PF4) antibodies generated during the type II heparin-induced thrombocytopenia (HIT type II) by ELISA.. Reason: Potential for defect for Asserachrom HPIA lot 271288, which can affect every well in the impacted strip and produce negative results.
What is being recalled
Asserachrom HPIA. Model/Catalog Number: 00615. The Asserachrom HPIA kit is intended for the qualitative detection in plasma or serum of the anti-heparin-platelet factor 4 (PF4) antibodies generated during the type II heparin-induced thrombocytopenia (HIT type II) by ELISA.
Why
Potential for defect for Asserachrom HPIA lot 271288, which can affect every well in the impacted strip and produce negative results.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
Worldwide - US Nationwide distribution in the states of AL, CA, FL, GA, HI, IA, MA, MI, NJ, NV, NY, OH, TN, WI and the countries of Belgium, Canada, Denmark, Algeria, Spain, Finland, France, Great Britain, Italy, Lebanon, Saudi Arabia, Uganda. (states named: Alabama, California, Florida, Georgia, Hawaii, Iowa, Massachusetts, Michigan, New Jersey, Nevada, New York, Ohio, Tennessee, Wisconsin)
How to tell if yours is affected
Reference Number: 00615. UDI: (01)03607450006155(11)240531(17)260531(10)271288(241)00615. Lot Number: 271288.
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-1084-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | December 15, 2025 |
| Reported | January 21, 2026 |
| Company | Diagnostica Stago, Inc. |
| Company location | Parsippany, NJ, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 438 units |