Recall Ledger

Abre Venous Self-Expanding Stent System recall (Medtronic Inc)

Medical deviceClass IIOngoing

Summary: Medtronic Inc is recalling Abre Venous Self-Expanding Stent System, Model numbers AB9G14120090 AB9G12100090 AB9G14080090 AB9G12080090 AB9U12060090 AB9U14100090 AB9U14080090 AB9U12080090 AB9U14060090 AB9G14100090 AB9G12060090. Reason: Stents may be impacted by a Nitinol anomaly which causes a reduced stent diameter when deployed and corresponding reduced radial force. Issue may result in stent migration.

What is being recalled

Abre Venous Self-Expanding Stent System, Model numbers AB9G14120090 AB9G12100090 AB9G14080090 AB9G12080090 AB9U12060090 AB9U14100090 AB9U14080090 AB9U12080090 AB9U14060090 AB9G14100090 AB9G12060090

Why

Stents may be impacted by a Nitinol anomaly which causes a reduced stent diameter when deployed and corresponding reduced radial force. Issue may result in stent migration.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Chile, Colombia, Costa Rica, Denmark, Dominican Republic, Finland, Germany, Jordan, Lebanon, Malaysia, Netherlands, New Zealand, Peru, Poland, Russia, Singapore, Spain, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, Uzbekistan, Vietnam.

How to tell if yours is affected

UDI-DI (GTIN): 00763000547332 00763000547271 00763000547318 00763000547264 00643169796225 00643169796300 00643169796294 00643169796232 00643169796287 00763000547325 00643169796300 00763000547257 Serial/Lot numbers C112511 C113253 C113795 C117416 C117482 C121084 C123953 C133327 C133329 C134478 C138060 C150325

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-1098-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedDecember 4, 2025
ReportedJanuary 21, 2026
CompanyMedtronic Inc
Company locationPlymouth, MN, United States
Initiated byVoluntary: Firm initiated
Quantity442

More recalls from Medtronic Inc · all medical device recalls · recalls from December 2025

Official FDA record