Recall Ledger

Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 10.1.1... recall (RAYSEARCH LABORATORIES AB)

Medical deviceClass IIOngoing

Summary: RAYSEARCH LABORATORIES AB is recalling Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 10.1.1 Software Version: RayStation 10B SP1 Product Description: Radiation Therapy Treatment Planning System Component: No. Reason: Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.

What is being recalled

Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 10.1.1 Software Version: RayStation 10B SP1 Product Description: Radiation Therapy Treatment Planning System Component: No

Why

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Refer to the attached documents: Distribution list 167168 US - updated, Distribution list 167168 OUS

How to tell if yours is affected

Lot Code: UDI: 0735000201047120220131 GTIN: 07350002010471 Serial Numbers: 10.1.1.54 Software Revision: RayStation 10B SP1 Expiration Date: 2027-02-01

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-1107-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedNovember 28, 2025
ReportedJanuary 28, 2026
CompanyRAYSEARCH LABORATORIES AB
Company locationStockholm, Sweden
Initiated byVoluntary: Firm initiated
Quantity0

More recalls from RAYSEARCH LABORATORIES AB · all medical device recalls · recalls from November 2025