Recall Ledger

Philips Azurion system recall (PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.)

Medical deviceClass IIOngoing

Summary: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. is recalling Philips Azurion system; Software Version Number: R3.1;. Reason: Philips has identified two (2) software issues affecting device systems that may result in loss of imaging (X-ray) functionality and/or loss of motorized movement, and/or incorrect image content and/or loss of data. Issue 1 - System remains in continuous restart mode after the start-up. Issue 2 - Longitudinal Position Error Applicable only to systems with Poly-G3 frontal stand.

What is being recalled

Philips Azurion system; Software Version Number: R3.1;

Why

Philips has identified two (2) software issues affecting device systems that may result in loss of imaging (X-ray) functionality and/or loss of motorized movement, and/or incorrect image content and/or loss of data. Issue 1 - System remains in continuous restart mode after the start-up. Issue 2 - Longitudinal Position Error Applicable only to systems with Poly-G3 frontal stand.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

International distribution in the countries of AE, AR, AT, AZ, BE, BG, BH, BR, BS, CH, CL, CO, CR, CY, CZ, DE, DK, DZ, EC, EG, ES, GB, GR, GT, ID, IE, IL, IN, IQ, IT, JE, JP, LB, MA, MO, NI, NL, NO, PE, PL, RO, SA, SE, SV, TR, UA, VE, VN, ZA; (states named: Arkansas, Arizona, Illinois, Indiana, Massachusetts, Missouri)

How to tell if yours is affected

Software Version Number: R3.1; (1) Model Number: 722221; System Product Name: Azurion 3 M12; UDI-DI: 00884838099203; Serial Numbers: (2) Model Number: 722222; System Product Name: Azurion 3 M15; UDI-DI: 00884838099210; Serial Numbers: (3) Model Number: 722223; System Product Name: Azurion 7 M12; UDI-DI: 00884838099241; Serial Numbers: (4) Model Number: 722224; System Product Name: Azurion 7 M20; UDI-DI: 00884838099258; Serial Numbers: (5) Model Number: 722225; System Product Name: Azurion 7 B12; UDI-DI: 00884838099265; Serial Numbers: (6) Model Number: 722226; System Product Name: Azurion 7 B20; UDI-DI: 00884838099272; Serial Numbers: (7) Model Number: 722227; System Product Name: Azurion 5 M12; UDI-DI: 00884838099227; Serial Numbers: (8) Model Number: 722228; System Product Name: Azurion 5 M20; UDI-DI: 00884838099234; Serial Numbers: (9) Model Number: 722229; System Product Name: Azurion 3 M12; UDI-DI: 00884838116726; Serial Numbers: 5, 9, 32, 40, 35, 31, 33; (10) Model Number: 722230; System Product Name: Azurion 3 M15; UDI-DI: 00884838116733; Serial Numbers: 25, 54, 50, 52, 55, 30, 63, 60, 67; (11) Model Number: 722231; System Product Name: Azurion 5 M12; UDI-DI: 00884838116740; Serial Numbers: 62, 7, 8, 14, 17, 63, 6, 66, 59, 69, 56, 67, 48, 68, 80, 93, 87, 94, 81, 92, 18; (12) Model Number: 722232; System Product Name: Azurion 5 M20; UDI-DI: 00884838116757; Serial Numbers: 97, 39, 31, 100, 104, 93, 115, 113, 118, 119, 89, 54, 67, 7, 128, 86, 110, 132, 130, 131, 123, 137, 139, 117, 260; (13) Model Number: 722233; System Product Name: Azurion 7 M12; UDI-DI: 00884838116764; Serial Numbers: 239, 100, 233, 240, 176, 227, 223, 130, 161, 194, 243, 56, 193, 280, 75, 253, 156, 288, 72, 281, 304, 300, 105, 295, 312, 310, 285, 292, 290, 298, 297; (14) Model Number: 722234; System Product Name: Azurion 7 M20; UDI-DI: 00884838116771; Serial Numbers: 731, 727, 261, 473, 683, 722, 562, 750, 745, 766, 265, 701, 400, 310, 421, 314, 182, 579, 96, 763, 272, 612, 15, 844, 105, 702, 751, 384, 851, 869, 743, 337, 651, 455, 706, 312, 563, 709, 764, 871, 846, 278, 713, 802, 868, 277, 742, 184, 819, 759, 910, 928, 415, 414, 945, 961, 355, 933, 692, 920, 569, 939, 909, 553, 837, 856, 571, 756, 941, 570, 761, 981, 353, 888, 964, 820, 746, 736, 876, 599, 1017, 908, 487, 983, 495; (15) Model Number: 722235; System Product Name: Azurion 7 B12; UDI-DI: 00884838116788; Serial Numbers: 96, 79, 111, 100, 101, 38, 27, 92, 78, 118, 48, 131, 135, 22, 125, 103, 138, 139, 137; (16) Model Number: 722236; System Product Name: Azurion 7 B20; UDI-DI: 00884838116801; Serial Numbers: 241, 297, 198, 215, 305, 274, 299, 78, 279, 287, 318, 353, 324, 331, 293, 245, 314, 310, 317, 312, 77, 358, 206, 383, 380, 370, 377, 385, 405, 379, 328;

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-1113-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedDecember 12, 2025
ReportedJanuary 28, 2026
CompanyPHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Company locationBest, Netherlands
Initiated byVoluntary: Firm initiated
Quantity228 units (all OUS)

More recalls from PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. · all medical device recalls · recalls from December 2025

Official FDA record