ICU Medical Plum Duo Infusion Pump recall (ICU Medical, Inc.)

Medical deviceClass IIOngoing

Summary: ICU Medical, Inc. is recalling ICU Medical Plum Duo Infusion Pump, List Numbers: 400020401 (US), 400021001 (Philippines). Reason: Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.

What is being recalled

ICU Medical Plum Duo Infusion Pump, List Numbers: 400020401 (US), 400021001 (Philippines)

Why

Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide distribution - US Nationwide and the country of Philippines.

How to tell if yours is affected

400020401 (US), Software Versions: 1.1.1, 1.1.2, 1.1.3, 1.1.4 400021001 (Philippines), Software Versions: 1.1.0

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-1118-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedDecember 15, 2025
ReportedJanuary 28, 2026
CompanyICU Medical, Inc.
Company locationLake Forest, IL, United States
Initiated byVoluntary: Firm initiated

More recalls from ICU Medical, Inc. · all medical device recalls · recalls from December 2025