TS-10/TS-10H Tube Sorter recall (Sysmex America, Inc.)
Medical deviceClass IIOngoing
Summary: Sysmex America, Inc. is recalling TS-10/TS-10H Tube Sorter, automatic sorting system for sorting barcoded sample tubes, Catalog Number BN168792, running software versions 00-17. Reason: Under specific conditions, the lift's electromagnetic holding force may disengage unexpectedly, allowing unintended downward movement of the lift.
What is being recalled
TS-10/TS-10H Tube Sorter, automatic sorting system for sorting barcoded sample tubes, Catalog Number BN168792, running software versions 00-17
Why
Under specific conditions, the lift's electromagnetic holding force may disengage unexpectedly, allowing unintended downward movement of the lift.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
US Nationwide.
How to tell if yours is affected
Serial Numbers: 11281 11611 11918 11790 12040 11991 12218 11950 11795 11533 11635 12054 11610 11453 11753 11539 11990 12298 11774 11541 11981 11788 11834 12005 11907 11524 12237 11246 12154 12104 12299 12176 11711 11420 12153 12177 12074 11405 11303 12039 11250 11336 12137 12075
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-1136-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | December 15, 2025 |
| Reported | January 28, 2026 |
| Company | Sysmex America, Inc. |
| Company location | Lincolnshire, IL, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 44 units |