Recall Ledger

Medline Kits containing Tego Connectors Medline DIALYSIS ON/OFF KIT SKU EBSI1746 recall (Medline Industries, LP)

Medical deviceClass IIOngoing

Summary: Medline Industries, LP is recalling Medline Kits containing Tego Connectors Medline DIALYSIS ON/OFF KIT SKU EBSI1746. Reason: Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.

What is being recalled

Medline Kits containing Tego Connectors Medline DIALYSIS ON/OFF KIT SKU EBSI1746

Why

Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

US Nationwide distribution.

How to tell if yours is affected

UDI-DI (ea) 10653160373582 UDI-DI (case) 00653160373585 Kit lots 2024051380 2024052080 2024052180 2024052280 2024071580 2025061790 2025072990

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-1464-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedJanuary 8, 2026
ReportedMarch 4, 2026
CompanyMedline Industries, LP
Company locationNorthfield, IL, United States
Initiated byVoluntary: Firm initiated
Quantity2020

More recalls from Medline Industries, LP · all medical device recalls · recalls from January 2026

Official FDA record