Owner's Booklets and Instructions for Use that are used with the following blood... recall (Trividia Health, Inc.)
Medical deviceClass IOngoing
Summary: Trividia Health, Inc. is recalling Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description/Item: Walgreens TRUE METRIX GO Kit/RF4001-01BK, CVS TRUE METRIX GO Kit/RF4007-01, Meijer TRUE METRIX GO Kit/RF4019-01, Fred's Phamarcy TRUE METRIX GO Kit/RF4023-01, HEB TRUE . Reason: The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.
What is being recalled
Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description/Item: Walgreens TRUE METRIX GO Kit/RF4001-01BK, CVS TRUE METRIX GO Kit/RF4007-01, Meijer TRUE METRIX GO Kit/RF4019-01, Fred's Phamarcy TRUE METRIX GO Kit/RF4023-01, HEB TRUE METRIX GO Kit/RF4031-01, Hyvee TRUE METRIX GO Kit/RF4048-01, Rite Aid TRUE METRIX GO Kit/RF4066-01, TopCo TRUE METRIX GO Kit/RF4209-01, TRUE METRIX GO Kit/RF4H01-01BK, TRUE METRIX GO NFRS Meter Only/RF4H01-40, TRUE METRIX GO Kit (Jamaica - mmol/L)/RF4i29-11BK, TRUE METRIX GO Kit (Australia - mmol/L)/RF4i81-11BK, TRUE METRIX GO Kit (United Kingdom - mmol/L)/RF4i82-11BK, TRUE METRIX GO Starter Kit (United Kingdom - mmol/L)/RF4i82-12BK
Why
The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.
How serious is it
Class I is the most serious level: FDA considers there is a reasonable probability that using or being exposed to this product will cause serious health problems or death.
Where it was sold
Worldwide distribution - US Nationwide including Puerto Rico and the countries of Mexico, United Kingdom, Jamaica, Trinidad & Tobago, and Australia. (states named: Puerto Rico)
How to tell if yours is affected
Device Item/UDI-DI: RF4001-01BK/311917178691, RF4007-01/50428560402, RF4019-01/708820721098, RF4023-01/21292009335, RF4031-01/41220136131, RF4048-01/75450843576, RF4066-01/11822002783, RF4209-01/36800234567, RF4H01-01BK/21292006112, RF4H01-40/21292009793. International: RF4i29-11BK/N/A, RF4i81-11BK/N/A, RF4i82-11BK/21292012229, RF4i82-12BK/21292009984. Owner's Booklet or IFU Description/Part and Revision: TRUE METRIX GO Owner's Booklet/ RF4TVH03 Rev 57 and prior, TRUE METRIX GO Owner's Booklet (English)/RF4TVH35 Rev 51 and prior, TRUE METRIX GO Owner's Booklet (Spanish)/RF4TVH35S Rev 51 and prior, TRUE METRIX GO Blood Glucose Monitoring System IFU Trividia House Brand/ RF4ITV16 Rev 55 and prior, TRUE METRIX GO System Instructions for Use (IFU) for Trividia Health UK Limited/ RF4UKT16 Rev 56 and prior, Trividia Health LATAM- TRUE METRIX GO System IFU/ RF4LAT16 Rev 51 and prior, TRUE METRIX GO System Instructions for Use (IFU) for Trividia Health Australia/ RF4AUT16 Rev 57 and prior. Device serial numbers: Prefix "F" followed by an eight-digit numeric sequence.
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-1548-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class I |
| Status | Ongoing |
| Recall started | February 6, 2026 |
| Reported | March 18, 2026 |
| Company | Trividia Health, Inc. |
| Company location | Fort Lauderdale, FL, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 1,407,465 |