BioFlo DuraMax Catheter recall (Merit Medical Systems, Inc.)
Medical deviceClass IOngoing
Summary: Merit Medical Systems, Inc. is recalling BioFlo DuraMax Catheter, REF: H965103028011/A, H965103028021/A, H965103028021/EU, H965103028031/A, H965103028031/EU, H965103028041/A, H965103028041/EU, H965103028051/A, H965103028051/EU, H965103028061/A, H965103028061/EU, H965103028070/A, H965103028070/EU, H965103028080/A, H965103028080/EU, H9651030. Reason: 16F dual-valved splittable sheath introducer due to a design defect, may not split as intended, which may result in hemorrhage, foreign bodies, procedure delay, embolization/thrombosis, impaired catheter function, loss of vessel for future vascular access.
What is being recalled
BioFlo DuraMax Catheter, REF: H965103028011/A, H965103028021/A, H965103028021/EU, H965103028031/A, H965103028031/EU, H965103028041/A, H965103028041/EU, H965103028051/A, H965103028051/EU, H965103028061/A, H965103028061/EU, H965103028070/A, H965103028070/EU, H965103028080/A, H965103028080/EU, H965103028090/A, H965103028090/EU, H965103028161/A, H965103028181/A, H965103028191/A, H965103028191/EU, H965103028201/A, H965103028211/A, H965103028220/A, H965103028230/A, H965103028240/A, H965103038181/A
Why
16F dual-valved splittable sheath introducer due to a design defect, may not split as intended, which may result in hemorrhage, foreign bodies, procedure delay, embolization/thrombosis, impaired catheter function, loss of vessel for future vascular access.
How serious is it
Class I is the most serious level: FDA considers there is a reasonable probability that using or being exposed to this product will cause serious health problems or death.
Where it was sold
Worldwide distribution: US (nationwide) to states of: TX, LA, OR, NM, MI, FL, CO, CA, GA, NC, AK, NY, PA, WI, NJ, VA, AZ, IN, KY, MO, DE, SC, MT, NE, WV, OH, MA, MN, CT, MD, KS, WA, TN, ID, UT, DC, MS, IL, OK, AL, RI, AR, NV and OUS (International) to countries of : Canada, Spain, United Arab Emirates, Brazil, Jordan, Iraq, France, Indonesia, Malaysia, Thailand, Slovakia, Colombia, Saudi Arabia, Belize, Hong Kong, Mexico, Austria, Belgium, Chile, Germany, Sweden, Ireland, Poland, Latvia, Italy, South Africa, Switzerland, Bulgaria, Peru, Argentina, Qatar, Greece, Georgia, United Kingdom, Brunei Darussalam, Ecuador, Panama, Netherlands, Venezuela, Bolivarian Republic of, Costa Rica, Viet Nam, Cayman Islands, Iran, Islamic Republic of, Australia, Nicaragua, Dominican Republic, Bolivia, Plurinational State of, El Salvador, Guatemala, Honduras, Japan (states named: Alaska, Alabama, Arkansas, Arizona, California, Colorado, Connecticut, District of Columbia, Delaware, Florida, Georgia, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Michigan, Minnesota, Missouri, Mississippi, Montana, North Carolina, Nebraska, New Jersey, New Mexico, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, Tennessee, Texas, Utah, Virginia, Washington, Wisconsin, West Virginia)
How to tell if yours is affected
REF:UDI-DI/Lot(Expiration): H965103028011/A:00884450793206/I3105589, I3115469, I3314948, I3343030; H965103028021/A:00884450793213/I2985084, I3235297, I3302100, I3343023; H965103028021/EU:00884450840733/I3254932; H965103028031/A:00884450786215/I2890967, I2909607, I2909610, I2909613, I2922255, I2928443, I2928445, I2973185, I2975817, I2979931, I2979932, I3010126, I3015730, I3015731, I3072562, I3108161, I3115487, I3120008, I3156395, I3161825, I3178917, I3202927, I3236144, I3236147, I3236157, I3236162, I3236164, I3368281; H965103028031/EU:00884450835326/I3219447, I3270828, I3288591; H965103028041/A:00884450786253/I2838431, I2910263, I2910264, I2910265, I2973186, I2975804, I2975805, I2979959, I2979960, I3015758, I3064923, I3064925, I3072563, I3108163, I3120003, I3120006, I3156399, I3165147, I3175748, I3178899, I3208677, I3236145, I3236148, I3236158, I3236163, I3236165, I3312321, I3315961, I3323271, I3347039; H965103028041/EU:00884450835333/I3217180, I3270829, I3302439; H965103028051/A:00884450793220/I2929337, I2997999, I3311219; H965103028051/EU:00884450840740/I3219448, I3270832, I3302440; H965103028061/A:00884450793237/I2920321, I2980492, I3154945, I3254911; H965103028061/EU:00884450835340/I3196140, I3270833, I3312332; H965103028070/A:00884450793244/I3015819, I3113237, I3170179, I3247943, I3288616; H965103028070/EU:00884450840757/I3255305, I3307483; H965103028080/A:00884450793251/I3002021, I306492, I3178919, I3302450, I3343031; H965103028080/EU:00884450840764/I3196137, I3302445, I3334625; H965103028090/A:00884450792353/I2920165, I3103138, I3178922, I3290565; H965103028090/EU:00884450840771/I3196135, I3302441, I3352493; H965103028161/A:00884450793268/I2914680, I2980457), I3098578; H965103028181/A:00884450793282/I2890978, I2902665, I2909617, I2909618, I2929338, I3076856, I3085067, I3098579, I3178915, I3254917, I3255301, I3283028; H965103028191/A:00884450793299/H2936911, I2870746, I2908636, I2909637, I2910257, I2920337, I2927199, I3144031, I3164176, I3219662, I3236146, I3288655, I3368110; H965103028191/EU:00884450835371/I3309111; H965103028201/A:00884450793305/I2973187, I3173621, I3178898), I3236149, I3288573; H965103028211/A:00884450793312/I2909764, I3144033, I3302435, I3343037; H965103028220/A:00884450793329/I2909788, I3394016; H965103028230/A:00884450793336/I2982337, I3044847, I3191920, I3288574, I3334623; H965103028240/A:00884450792360/I2929339, I3002035, I3049565, I3108165, I3254923, I3302451; H965103038181/A:00884450786314/I2890952, I2902661, I2902662
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-1579-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class I |
| Status | Ongoing |
| Recall started | February 13, 2026 |
| Reported | April 1, 2026 |
| Company | Merit Medical Systems, Inc. |
| Company location | South Jordan, UT, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 47,153 |
More recalls from Merit Medical Systems, Inc. · all medical device recalls · recalls from February 2026