Custom Abutment Ti recall (Straumann USA LLC)
Medical deviceClass IIOngoing
Summary: Straumann USA LLC is recalling Custom Abutment Ti, for Straumann RC, incl. TAN screw; Catalog Number/Art.: 027.4620;. Reason: A material mix-up had occurred during the manufacturing of Straumann¿ Custom Abutments. More specifically, a raw material bar of TAN-material (SAP #701924 / Lot 7130439) was incorrectly identified as a Titanium Bar Ti Gr4 D (SAP #701922 / Lot 7130430). As a result, RC Titanium abutments were manufactured using the TAN Bar material on Mill-from-Bar Machine 396-06.
What is being recalled
Custom Abutment Ti, for Straumann RC, incl. TAN screw; Catalog Number/Art.: 027.4620;
Why
A material mix-up had occurred during the manufacturing of Straumann¿ Custom Abutments. More specifically, a raw material bar of TAN-material (SAP #701924 / Lot 7130439) was incorrectly identified as a Titanium Bar Ti Gr4 D (SAP #701922 / Lot 7130430). As a result, RC Titanium abutments were manufactured using the TAN Bar material on Mill-from-Bar Machine 396-06.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
Worldwide - US Nationwide distribution in the states of CA, CO, FL, IA, IL, MN, MO, MT, OR, WA; and the country of Canada. (states named: California, Colorado, Florida, Iowa, Illinois, Minnesota, Missouri, Montana, Oregon, Washington)
How to tell if yours is affected
Catalog Number/Art.: 027.4620; UDI-DI: 07630031713766; Order number (lot): 92465US_12039_00_b00 , 92465US_12040_00_b00 , 92465US_12047_00_b00 , 92465US_12048_00_b00 , 70959US_02227_00_b00 , 94021US_02348_00_b00, 70416US_07543_00_b00 , 91395US_01687_00_b00 , 91545US_03284_00_b00 , 91545US_03285_00_b00 , 71711US_00420_00_b00 , 91853US_06313_00_b00 , 72747US_00247_00_b00;
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-1632-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | March 17, 2026 |
| Reported | April 1, 2026 |
| Company | Straumann USA LLC |
| Company location | Andover, MA, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 13 units |