Recall Ledger

Custom Abutment AS Ti recall (Straumann USA LLC)

Medical deviceClass IIOngoing

Summary: Straumann USA LLC is recalling Custom Abutment AS Ti, for Straumann RC, incl. TAN screw; Catalog Number/Art.: 010.5438;. Reason: A material mix-up had occurred during the manufacturing of Straumann¿ Custom Abutments. More specifically, a raw material bar of TAN-material (SAP #701924 / Lot 7130439) was incorrectly identified as a Titanium Bar Ti Gr4 D (SAP #701922 / Lot 7130430). As a result, RC Titanium abutments were manufactured using the TAN Bar material on Mill-from-Bar Machine 396-06.

What is being recalled

Custom Abutment AS Ti, for Straumann RC, incl. TAN screw; Catalog Number/Art.: 010.5438;

Why

A material mix-up had occurred during the manufacturing of Straumann¿ Custom Abutments. More specifically, a raw material bar of TAN-material (SAP #701924 / Lot 7130439) was incorrectly identified as a Titanium Bar Ti Gr4 D (SAP #701922 / Lot 7130430). As a result, RC Titanium abutments were manufactured using the TAN Bar material on Mill-from-Bar Machine 396-06.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide - US Nationwide distribution in the states of CA, CO, FL, IA, IL, MN, MO, MT, OR, WA; and the country of Canada. (states named: California, Colorado, Florida, Iowa, Illinois, Minnesota, Missouri, Montana, Oregon, Washington)

How to tell if yours is affected

Catalog Number/Art.: 010.5438; UDI-DI: 07630031773937; Order number (lot): 71315US_00729_01_b00, 91076US_03269_00_b00, 72509US_01048_00_b00, 72839US_00065_00_b00, 39133US_11508_02_b00, 4202CA_00558_00_b00, 72630CA_00109_00_b00;

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-1633-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedMarch 17, 2026
ReportedApril 1, 2026
CompanyStraumann USA LLC
Company locationAndover, MA, United States
Initiated byVoluntary: Firm initiated
Quantity7 units

More recalls from Straumann USA LLC · all medical device recalls · recalls from March 2026