Brand Name: TECNIS Eyhance IOL with TECNIS Simplicity" Delivery System Product Name:... recall (AMO Puerto Rico Manufacturing, Inc.)
Medical deviceClass IIOngoing
Summary: AMO Puerto Rico Manufacturing, Inc. is recalling Brand Name: TECNIS Eyhance IOL with TECNIS Simplicity" Delivery System Product Name: TECNIS Eyhance IOL with TECNIS Simplicity" Delivery System Model/Catalog Number: DIB00 Software Version: N/A Product Description: The TECNIS SIMPLICITY" Delivery System is designed to provide a sterile,controlle. Reason: Due to identifying a limited quantity of intraocular lenses that may have a haptic that sticks to the optic which prevents the lens from unfolding as consistently as expected.
What is being recalled
Brand Name: TECNIS Eyhance IOL with TECNIS Simplicity" Delivery System Product Name: TECNIS Eyhance IOL with TECNIS Simplicity" Delivery System Model/Catalog Number: DIB00 Software Version: N/A Product Description: The TECNIS SIMPLICITY" Delivery System is designed to provide a sterile,controlled and touch-free method of delivering the lens into the eye. The lens is preloaded and pre-assembled in the delivery system. The TECNIS SIMPLICITY" Delivery System Model DIB00 contains the TECNISEyhance" IOL, which is a one-piece, foldable, posterior chamber lens with an overall diameter of 13.0 mm and an optic diameter of 6.0 mm. The TECNIS SIMPLICITY" Delivery System is designed to provide a sterile,controlled and touch-free method of delivering the lens into the eye. Component: No
Why
Due to identifying a limited quantity of intraocular lenses that may have a haptic that sticks to the optic which prevents the lens from unfolding as consistently as expected.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
U.S Nationwide distribution in the states of CA,FL, IL, ME,MS, NJ, NY, OK, OR, TX, and WI. (states named: California, Florida, Illinois, Maine, Mississippi, New Jersey, New York, Oklahoma, Oregon, Texas, Wisconsin)
How to tell if yours is affected
Lot Code: GTIN: 05050474731776 SKU: DIB00U0210-12 Serial Numbers: 2409342403 2410072403 2409612403 2409692403 2409752403 2891682403 2891132403 2892032403 2531882403 2531902403 2531212403 2530842403 2531242403 2531892403 2531622403
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-1634-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | February 19, 2026 |
| Reported | April 1, 2026 |
| Company | AMO Puerto Rico Manufacturing, Inc. |
| Company location | Anasco, PR, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 361 |