Recall Ledger

ARTIS Pheno VE30A and VE40A recall (Siemens Medical Solutions USA, Inc)

Medical deviceClass IIOngoing

Summary: Siemens Medical Solutions USA, Inc is recalling ARTIS Pheno VE30A and VE40A, Model 10849000. Reason: During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient.

What is being recalled

ARTIS Pheno VE30A and VE40A, Model 10849000

Why

During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

U.S. and O.U.S.

How to tell if yours is affected

UDI 04056869046877

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-1814-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedMarch 12, 2026
ReportedApril 22, 2026
CompanySiemens Medical Solutions USA, Inc
Company locationMalvern, PA, United States
Initiated byFDA Mandated
Quantity73

More recalls from Siemens Medical Solutions USA, Inc · all medical device recalls · recalls from March 2026

Official FDA record