Recall Ledger

Sequencing Agent SEQ0067 recall (Foundation Medicine, Inc.)

Medical deviceClass IIOngoing

Summary: Foundation Medicine, Inc. is recalling Sequencing Agent SEQ0067, component of FoundationOne CDx (F1CDx). Reason: Possible increase in phasing, potentially resulting in false-positive BARD1 splice site indel artifact variants.

What is being recalled

Sequencing Agent SEQ0067, component of FoundationOne CDx (F1CDx)

Why

Possible increase in phasing, potentially resulting in false-positive BARD1 splice site indel artifact variants

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

US Nationwide distribution in the states of Massachusetts and North Carolina. (states named: Massachusetts, North Carolina)

How to tell if yours is affected

Model Number: SEQ0067. Lot Numbers: 24346, 24386, 24488.

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-1927-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedMarch 20, 2026
ReportedApril 29, 2026
CompanyFoundation Medicine, Inc.
Company locationCambridge, MA, United States
Initiated byVoluntary: Firm initiated
Quantity163 units

More recalls from Foundation Medicine, Inc. · all medical device recalls · recalls from March 2026

Official FDA record