Recall Ledger

iLet Bionic Pancreas recall (Beta Bionics, Inc.)

Medical deviceClass IIOngoing

Summary: Beta Bionics, Inc. is recalling iLet Bionic Pancreas, REF: BB1001. Reason: insulin pump software versions 1.4.2, and 1.4.3 issue with Dexcom Continuous Glucose Monitoring (CGM) G7 sensor may result in delayed CGM reading then switch to non-CGM dosing, leading to only basal/requested meal doses provided, no correction doses/insulin reductions/suspensions for high/low glucose unless blood glucose entered for sensor life or replacement. Potential health risks to users include severe hyperglycemia with and without DKA and severe hypoglycemia.

What is being recalled

iLet Bionic Pancreas, REF: BB1001

Why

insulin pump software versions 1.4.2, and 1.4.3 issue with Dexcom Continuous Glucose Monitoring (CGM) G7 sensor may result in delayed CGM reading then switch to non-CGM dosing, leading to only basal/requested meal doses provided, no correction doses/insulin reductions/suspensions for high/low glucose unless blood glucose entered for sensor life or replacement. Potential health risks to users include severe hyperglycemia with and without DKA and severe hypoglycemia.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

US Nationwide distribution in the states of TX, PA, AL, MI, OH, FL, TN, GA. (states named: Alabama, Florida, Georgia, Michigan, Ohio, Pennsylvania, Tennessee, Texas)

How to tell if yours is affected

Software versions 1.4.3, and 1.4.2

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-1996-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedMarch 12, 2026
ReportedMay 6, 2026
CompanyBeta Bionics, Inc.
Company locationIrvine, CA, United States
Initiated byVoluntary: Firm initiated
Quantity1,080

More recalls from Beta Bionics, Inc. · all medical device recalls · recalls from March 2026