Medline Convenience Kits: 1) LITHOTOMY PACK recall (Medline Industries, LP)
Medical deviceClass IIOngoing
Summary: Medline Industries, LP is recalling Medline Convenience Kits: 1) LITHOTOMY PACK, Model Number: DYNJ83185; 2) LITHOTOMY PACK, Model Number: DYNJ83185A; 3) LAP BASIN, Model Number: DYNJVB91109D. Reason: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
What is being recalled
Medline Convenience Kits: 1) LITHOTOMY PACK, Model Number: DYNJ83185; 2) LITHOTOMY PACK, Model Number: DYNJ83185A; 3) LAP BASIN, Model Number: DYNJVB91109D
Why
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
Worldwide - US Nationwide distribution.
How to tell if yours is affected
1) DYNJ83185, UDI-DI: 10195327247058(each), 40195327247059(case), Lot Number: 22LDA486; 2) DYNJ83185, UDI-DI: 10195327247058(each), 40195327247059(case), Lot Number: 23CDA474; 3) DYNJ83185, UDI-DI: 10195327247058(each), 40195327247059(case), Lot Number: 23CDB552; 4) DYNJ83185, UDI-DI: 10195327247058(each), 40195327247059(case), Lot Number: 23FDC392; 5) DYNJ83185A, UDI-DI: 10195327468606(each), 40195327468607(case), Lot Number: 23HDB637; 6) DYNJ83185A, UDI-DI: 10195327468606(each), 40195327468607(case), Lot Number: 24CDA984; 7) DYNJ83185A, UDI-DI: 10195327468606(each), 40195327468607(case), Lot Number: 24CDB327; 8) DYNJ83185A, UDI-DI: 10195327468606(each), 40195327468607(case), Lot Number: 24CDB829; 9) DYNJ83185A, UDI-DI: 10195327468606(each), 40195327468607(case), Lot Number: 24DDB769; 10) DYNJ83185A, UDI-DI: 10195327468606(each), 40195327468607(case), Lot Number: 24EDB468; 11) DYNJ83185A, UDI-DI: 10195327468606(each), 40195327468607(case), Lot Number: 24FDB302; 12) DYNJ83185A, UDI-DI: 10195327468606(each), 40195327468607(case), Lot Number: 24GDB922; 13) DYNJ83185A, UDI-DI: 10195327468606(each), 40195327468607(case), Lot Number: 24IDB525; 14) DYNJ83185A, UDI-DI: 10195327468606(each), 40195327468607(case), Lot Number: 24KDB955; 15) DYNJVB91109D, UDI-DI: 10198459514838(each), 40198459514839(case), Lot Number: 25LLA489
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-2100-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | January 7, 2026 |
| Reported | May 13, 2026 |
| Company | Medline Industries, LP |
| Company location | Northfield, IL, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 826 units |
More recalls from Medline Industries, LP · all medical device recalls · recalls from January 2026