Medline Convenience Kits: 1) CENTRAL LINE UNIVERSAL INSERTI recall (Medline Industries, LP)

Medical deviceClass IIOngoing

Summary: Medline Industries, LP is recalling Medline Convenience Kits: 1) CENTRAL LINE UNIVERSAL INSERTI, Model Number: CVI3780; 2) LVAD DRIVELINE TRAY, Model Number: DM1035; 3) KIT SURG ONC INSERT CENT.VENOU, Model Number: DYKMBNDL43. Reason: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

What is being recalled

Medline Convenience Kits: 1) CENTRAL LINE UNIVERSAL INSERTI, Model Number: CVI3780; 2) LVAD DRIVELINE TRAY, Model Number: DM1035; 3) KIT SURG ONC INSERT CENT.VENOU, Model Number: DYKMBNDL43

Why

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide - US Nationwide distribution.

How to tell if yours is affected

1) CVI3780, UDI-DI: 10193489434866(each), 40193489434867(case), Lot Number: 23KBO637; 2) CVI3780, UDI-DI: 10193489434866(each), 40193489434867(case), Lot Number: 24ABC276; 3) DM1035, UDI-DI: 10193489442755(each), 40193489442756(case), Lot Number: 21BBN221; 4) DYKMBNDL43, UDI-DI: 10193489478891(each), 40193489478892(case), Lot Number: 21ABJ003; 5) DYKMBNDL43, UDI-DI: 10193489478891(each), 40193489478892(case), Lot Number: 21ABO838; 6) DYKMBNDL43, UDI-DI: 10193489478891(each), 40193489478892(case), Lot Number: 21DBD662; 7) DYKMBNDL43, UDI-DI: 10193489478891(each), 40193489478892(case), Lot Number: 21IBJ337; 8) DYKMBNDL43, UDI-DI: 10193489478891(each), 40193489478892(case), Lot Number: 21JBJ253; 9) DYKMBNDL43, UDI-DI: 10193489478891(each), 40193489478892(case), Lot Number: 22CBU342; 10) DYKMBNDL43, UDI-DI: 10193489478891(each), 40193489478892(case), Lot Number: 22EBL039; 11) DYKMBNDL43, UDI-DI: 10193489478891(each), 40193489478892(case), Lot Number: 22GBF510; 12) DYKMBNDL43, UDI-DI: 10193489478891(each), 40193489478892(case), Lot Number: 22GBN333

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2103-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedJanuary 7, 2026
ReportedMay 13, 2026
CompanyMedline Industries, LP
Company locationNorthfield, IL, United States
Initiated byVoluntary: Firm initiated
Quantity1056 units