Integris-Allura system recall (PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.)

Medical deviceClass IIOngoing

Summary: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. is recalling Integris-Allura system; System Code Description (Model Numbers): Integris CV Cesar-Powerpack-Visub-Nicol (722030), Integris Allura 15 & 12 (monoplane) (722043), INTEGRIS Allura 15-12 (biplane) (722044), Integris-Allura 9 (722018), Integris-Allura 9 (Biplane) (722021), Integris CV Cesar-Powerpack (72. Reason: The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.

What is being recalled

Integris-Allura system; System Code Description (Model Numbers): Integris CV Cesar-Powerpack-Visub-Nicol (722030), Integris Allura 15 & 12 (monoplane) (722043), INTEGRIS Allura 15-12 (biplane) (722044), Integris-Allura 9 (722018), Integris-Allura 9 (Biplane) (722021), Integris CV Cesar-Powerpack (722030), Integris H5000 C / Allura 9C (722016);

Why

The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Domestic: Nationwide; International: Canada, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Ethiopia, Finland, France, French Polynesia, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Republic of, Kosovo, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Malaysia, Maldives, Malta, Martinique, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Viet Nam, Yemen; (states named: Georgia)

How to tell if yours is affected

1. Integris CV Cesar-Powerpack-Visub-Nicol; Model: 722030; UDI-DI: N/A; Serial Numbers: 4888270, 59, 0, 33; 2. Integris Allura 15 & 12 (monoplane) (722043); UDI-DI: N/A; Serial Numbers: 365, 18151, 176, 239, 407, 426, 0; 3. INTEGRIS Allura 15-12 (biplane) (722044); UDI-DI: N/A; Serial Numbers: 86322, 30, 66; 4. Integris-Allura 9 (722018); UDI-DI: N/A; Serial Numbers: 26, 85; 5. Integris-Allura 9 (Biplane) (722021); UDI-DI: N/A; Serial numbers: 22, 75, 25, 108; 6. Integris H5000 C / Allura 9C (722016); UDI-DI: N/A; Serial Numbers: 4059001008, 4807240, S01H003553/000001;

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2229-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedApril 28, 2026
ReportedMay 27, 2026
CompanyPHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Company locationBest, Netherlands
Initiated byVoluntary: Firm initiated
Quantity25 units (11 US, 14 OUS)