Brand Name: Exacta Mix Product Name: Vented Micro-Volume Inlet Model/Catalog Number:... recall (Baxter Healthcare Corporation)
Medical deviceClass IIOngoing
Summary: Baxter Healthcare Corporation is recalling Brand Name: Exacta Mix Product Name: Vented Micro-Volume Inlet Model/Catalog Number: H938175 Product Description: Tubing sets with push-on connector on one end (to attach to the Fluid Selector Valve) and a spike or other appropriate connector on the opposite end to attach the source container.. Reason: Affected lots may contain incorrect 0.8-micron filters. Use of a 0.8-micron filter in place of the validated 3-micron filter may lead to reduced flow efficiency, incomplete or slower transfers of ingredients, or system alarms. This issue may lead to clinical effects such as electrolyte imbalance, metabolic instability, delayed recovery, etc.
What is being recalled
Brand Name: Exacta Mix Product Name: Vented Micro-Volume Inlet Model/Catalog Number: H938175 Product Description: Tubing sets with push-on connector on one end (to attach to the Fluid Selector Valve) and a spike or other appropriate connector on the opposite end to attach the source container.
Why
Affected lots may contain incorrect 0.8-micron filters. Use of a 0.8-micron filter in place of the validated 3-micron filter may lead to reduced flow efficiency, incomplete or slower transfers of ingredients, or system alarms. This issue may lead to clinical effects such as electrolyte imbalance, metabolic instability, delayed recovery, etc.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
Worldwide - US Nationwide distribution including in the states of Alabama, Arkansas, Arizona, California, Colorado, Florida, Georgia, Hawaii, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maine, Maryland, Michigan, Minnesota, Missouri, Mississippi, Montana, North Carolina, North Dakota, Nebraska, New Hampshire, New Jersey, New Mexico, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, South Dakota, Tennessee, Texas, Utah, Virginia, Washington, Wisconsin, and West Virginia. The countries of Taiwan and Uruguay. (states named: Alabama, Arkansas, Arizona, California, Colorado, Florida, Georgia, Hawaii, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Maine, Michigan, Minnesota, Missouri, Mississippi, Montana, North Carolina, North Dakota, Nebraska, New Hampshire, New Jersey, New Mexico, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, South Dakota, Tennessee, Texas, Utah, Virginia, Washington, Wisconsin, West Virginia)
How to tell if yours is affected
Lot Code: Lot numbers: 804084, 804088, and 804089 UDI Number: 00085412475806
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-2262-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | April 22, 2026 |
| Reported | June 10, 2026 |
| Company | Baxter Healthcare Corporation |
| Company location | Deerfield, IL, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 105675 |
More recalls from Baxter Healthcare Corporation · all medical device recalls · recalls from April 2026