PhaseOne Antimicrobial Solution recall (Oculus Technologies of Mexico, S.A. de C.V.)
Medical deviceClass IIOngoing
Summary: Oculus Technologies of Mexico, S.A. de C.V. is recalling PhaseOne Antimicrobial Solution, Model/Catalog Number: 15050. Reason: The recall is being conducted due to corrosion of the internal metallic spring within the sprayer component. This condition is caused by material incompatibility between the product solution and the stainless steel spring and may result in visible discoloration and the presence of trace levels of metallic elements (including iron, chromium, and nickel) in the solution during use.
What is being recalled
PhaseOne Antimicrobial Solution, Model/Catalog Number: 15050
Why
The recall is being conducted due to corrosion of the internal metallic spring within the sprayer component. This condition is caused by material incompatibility between the product solution and the stainless steel spring and may result in visible discoloration and the presence of trace levels of metallic elements (including iron, chromium, and nickel) in the solution during use.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
US Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, IN, KY, MI, MS, NC, NJ, NY, TN, TX, UT, VA. (states named: Arizona, California, Colorado, Florida, Georgia, Illinois, Indiana, Kentucky, Michigan, Mississippi, North Carolina, New Jersey, New York, Tennessee, Texas, Utah, Virginia)
How to tell if yours is affected
UDI:00860009268616 Lot: 25G263
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-2275-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | April 27, 2026 |
| Reported | June 10, 2026 |
| Company | Oculus Technologies of Mexico, S.A. de C.V. |
| Company location | Zapopan, Mexico |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 5904 units |