Sharpoint 2.2MM recall (SSC- Surgical Specialties Corporation)

Medical deviceClass IIOngoing

Summary: SSC- Surgical Specialties Corporation is recalling Sharpoint 2.2MM, ANGLED, DOUBLE- BEVEL Model/Catalog Number: 72-2231 Microsurgical Knife. Reason: 3.0mm surgical knives labeled and packaged as 2.2mm surgical knives.

What is being recalled

Sharpoint 2.2MM, ANGLED, DOUBLE- BEVEL Model/Catalog Number: 72-2231 Microsurgical Knife

Why

3.0mm surgical knives labeled and packaged as 2.2mm surgical knives.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide - US Nationwide distribution in the states of Texas, Utah, Michigan and the countries of Turkey, Germany, Romania, Poland, United Kingdom. (states named: Michigan, Texas, Utah)

How to tell if yours is affected

UDI-DI 00848782027514; Lot Number FA22CHQ

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2309-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedApril 23, 2026
ReportedJune 10, 2026
CompanySSC- Surgical Specialties Corporation
Company locationChula Vista, CA, United States
Initiated byVoluntary: Firm initiated
Quantity1818 units