Allia IGS 3 Pulse angiographic X-ray system recall (GE Medical Systems, LLC)

Medical deviceClass IIOngoing

Summary: GE Medical Systems, LLC is recalling Allia IGS 3 Pulse angiographic X-ray system. Reason: GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.

What is being recalled

Allia IGS 3 Pulse angiographic X-ray system

Why

GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide - US Nationwide distribution.

How to tell if yours is affected

UDI-DI: 00195278719577; Serial Numbers: M3-24-053, M3-24-124, M3-25-029, M3-25-051, M3-25-053, M3-25-091, M3-25-098, M3-25-124, M3-26-010, M3-26-030

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2314-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedApril 20, 2026
ReportedJune 17, 2026
CompanyGE Medical Systems, LLC
Company locationWaukesha, WI, United States
Initiated byVoluntary: Firm initiated
Quantity9 units