Medline CCI Nerve Block Tray DYNJRA2742 Medline CPNB DYNJRA2716 recall (Medline Industries, LP)
Medical deviceClass IIOngoing
Summary: Medline Industries, LP is recalling Medline CCI Nerve Block Tray DYNJRA2742 Medline CPNB DYNJRA2716. Reason: Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
What is being recalled
Medline CCI Nerve Block Tray DYNJRA2742 Medline CPNB DYNJRA2716
Why
Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
US Nationwide distribution. OUS distribution pending.
How to tell if yours is affected
DYNJRA2742 UDI-DI 10198459627330 Lot 26AMA185 DYNJRA2716 UDI-DI 10198459567988 Lot 25KBA971
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-2415-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | April 27, 2026 |
| Reported | June 24, 2026 |
| Company | Medline Industries, LP |
| Company location | Northfield, IL, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 49,654 kits total |
More recalls from Medline Industries, LP · all medical device recalls · recalls from April 2026