Summary: AMD Medicom Inc. is recalling Ritmed Neurological Sponges, Description/REF: 0.5x0.5in, Non-Sterile/55992C; 0.5x1.0in, Non-Sterile/55993C; 0.5x3.0in, Non-Sterile/55994C; 0.75x0.75in, Non-Sterile/55995C; 1.0x1.0in, Non-Sterile/55996C; 1.0x3.0in, Non-Sterile/55997C; 0.5x1.5in, Non-Sterile/55998C. Reason: Neurological sponges used to moisten or absorb biological fluid during neurological surgery and may also be used to protect the brain during incision, may not consistently meet endotoxin specification for devices that contact the cerebrospinal fluid. Elevated endotoxin levels may lead to neurovascular adverse events, febrile response, and/or local transient inflammation, hypotension, or nausea.

What is being recalled

Ritmed Neurological Sponges, Description/REF: 0.5x0.5in, Non-Sterile/55992C; 0.5x1.0in, Non-Sterile/55993C; 0.5x3.0in, Non-Sterile/55994C; 0.75x0.75in, Non-Sterile/55995C; 1.0x1.0in, Non-Sterile/55996C; 1.0x3.0in, Non-Sterile/55997C; 0.5x1.5in, Non-Sterile/55998C

Why

Neurological sponges used to moisten or absorb biological fluid during neurological surgery and may also be used to protect the brain during incision, may not consistently meet endotoxin specification for devices that contact the cerebrospinal fluid. Elevated endotoxin levels may lead to neurovascular adverse events, febrile response, and/or local transient inflammation, hypotension, or nausea.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

US: MT, NC, VA, CA, TX, OH, TN, IL, RI (states named: California, Illinois, Montana, North Carolina, Ohio, Rhode Island, Tennessee, Texas, Virginia)

How to tell if yours is affected

UDI-DI: 20686864043205. Distributed After May 1st, 2022, including the following lots: 19316, 16409, 15463-7, 15463-6, 15463-5, 15463-4, 15463-3, 15463-2, 15463-1, 15463, 15192-1, 15192, 14813, 14039, 13673, 13289, 12641, 12206, 11851, 11020-1, 11020, 10586, 10392, 9243, 8160, 7784, 7748, 7044

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2458-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedMay 8, 2026
ReportedJune 24, 2026
CompanyAMD Medicom Inc.
Company locationMontreal, Canada
Initiated byVoluntary: Firm initiated
Quantity7,838,200