INZONE DETACHMENT SYSTEM recall (Stryker Neurovascular)
Medical deviceClass IIOngoing
Summary: Stryker Neurovascular is recalling INZONE DETACHMENT SYSTEM, REF: M00345100950. Reason: Embolization coil detachment system devices may experience premature battery drain causing devices to: 1) Not power on; 2) Power on with faint audible and visual indicators; 3) Be unable to detach a coil as intended, which may necessitate medical intervention if the procedure must be completed with an alternative technique.
What is being recalled
INZONE DETACHMENT SYSTEM, REF: M00345100950
Why
Embolization coil detachment system devices may experience premature battery drain causing devices to: 1) Not power on; 2) Power on with faint audible and visual indicators; 3) Be unable to detach a coil as intended, which may necessitate medical intervention if the procedure must be completed with an alternative technique.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
Worldwide - US Nationwide distribution including in the states of PA, AZ, FL, CA, IL, AK, NY, MI, NJ, WI, SD, KY, AR, TX, AL, TN, MO, MA, IN, NE, WV, MT, CT, NC, VA, NV, CO, DC, DE, OH, ME, GA, WA, SC, GU, MN, OK, NH, MD, ND, UT, HI, OR, LA, MS, KS, NM, ID, IA, VT and the countries of ARGENTINA, AUSTRIA, BELGIUM, BRASIL, BULGARIA, CANADA, CANARY ISLANDS, CHILE, COLOMBIA, CROATIA, CZECH REPUBLIC, DENMARK, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, HUNGARY, ICELAND, IRAN, IRELAND, ISRAEL, ITALIA, JAPAN, KOREA, LATVIA, LUXEMBOURG, MALAYSIA, MEXICO, NETHERLANDS, NORTHERN IRELAND, NORWAY, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH PACIFIC, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM. (states named: Alaska, Alabama, Arkansas, Arizona, California, Colorado, Connecticut, District of Columbia, Delaware, Florida, Georgia, Hawaii, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Maine, Michigan, Minnesota, Missouri, Mississippi, Montana, North Carolina, North Dakota, Nebraska, New Hampshire, New Jersey, New Mexico, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Vermont, Washington, Wisconsin, West Virginia)
How to tell if yours is affected
UDI-DI: 04546540697950. Lots: WMP133638, WMP133871, WMP133872, WMP133988, WMP133989, WMP134016, WMP134132, WMP134133, WMP134234, WMP134235, WMP134255, WMP134256, WMP134380, WMP134381, WMP134432, WMP134433, WMP134518, WMP134519, WMP134609, WMP134610, WMP134794, WMP134795, WMP134939, WMP134940, WMP135187, WMP135243, WMP135302, WMP135314, WMP135354, WMP135355, WMP135387, WMP135417, WMP135473, WMP135475, WMP135491, WMP135492, WMP135623, WMP135720, WMP135727, WMP135807, WMP135865, WMP135897, WMP136084, WMP136107, WMP136190, WMP136200, WMP136611, WMP136642, WMP136736, WMP136748, WMP136852, WMP136894, WMP136985
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-2552-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | April 6, 2026 |
| Reported | July 1, 2026 |
| Company | Stryker Neurovascular |
| Company location | Fremont, CA, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 44,937 |