Recall Ledger

Reprocessed Electrophysiology Catheters and Ultrasound Catheters Item no. 401206RH recall (Medline Industries, LP)

Medical deviceClass IOngoing

Summary: Medline Industries, LP is recalling Reprocessed Electrophysiology Catheters and Ultrasound Catheters Item no. 401206RH, St. Jude Medical Response" Diagnostic Electrophysiology Catheter 4 Elec. 10mm; Item no. 401305RH, St. Jude Medical Response" Diagnostic Electrophysiology Catheter 10 Elec. 2-5-2mm.. Reason: Expansion of RES 98277. Specific lots of reprocessed catheters may contain small particles of residual material. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection. If particulate were to dislodge during use and contact circulating blood or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.

What is being recalled

Reprocessed Electrophysiology Catheters and Ultrasound Catheters Item no. 401206RH, St. Jude Medical Response" Diagnostic Electrophysiology Catheter 4 Elec. 10mm; Item no. 401305RH, St. Jude Medical Response" Diagnostic Electrophysiology Catheter 10 Elec. 2-5-2mm.

Why

Expansion of RES 98277. Specific lots of reprocessed catheters may contain small particles of residual material. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection. If particulate were to dislodge during use and contact circulating blood or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.

How serious is it

Class I is the most serious level: FDA considers there is a reasonable probability that using or being exposed to this product will cause serious health problems or death.

Where it was sold

US Nationwide distribution in the state of TX. (states named: Texas)

How to tell if yours is affected

Lot EP250203 Item no. 401206RH - UDI-DI 10197344043767; Item no. 401305RH - UDI-DI 10197344043637.

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2615-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass I
StatusOngoing
Recall startedMay 15, 2026
ReportedJuly 8, 2026
CompanyMedline Industries, LP
Company locationNorthfield, IL, United States
Initiated byVoluntary: Firm initiated
Quantity3