Recall Ledger

MiniMed Go App recall (Medtronic MiniMed, Inc.)

Medical deviceClass IIOngoing

Summary: Medtronic MiniMed, Inc. is recalling MiniMed Go App, Model: MMT-8600 (iOS), MMT-8601 (Android). Reason: For MiniMed Go app version 1.0.0 and 1.1.0, when the Set day and night is turned OFF and missed dose or Correct high glucose alerts are enabled, the associated alerts will not be triggered during the nighttime. This could result in missed rapid-acting insulin doses without the user s knowledge, and may lead to delayed therapy and an increased risk of hyperglycemia.

What is being recalled

MiniMed Go App, Model: MMT-8600 (iOS), MMT-8601 (Android)

Why

For MiniMed Go app version 1.0.0 and 1.1.0, when the Set day and night is turned OFF and missed dose or Correct high glucose alerts are enabled, the associated alerts will not be triggered during the nighttime. This could result in missed rapid-acting insulin doses without the user s knowledge, and may lead to delayed therapy and an increased risk of hyperglycemia.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide - US Nationwide distribution in the states and territories of PR, OH, WA, IL, NC, PA, VA, MN, ID, CA, UT, MI, NJ, MS, SC, CO, AZ, OR, AR, MO, TX, AL, FL, NY, IN, TN, IA, WI, MA, LA, OK, GA, KS, NM, CT, KY, NH, NV, MD and the countries of Australia, Switzerland, Germany, Denmark, Spain, Finland, Iceland, Norway, Sweden. (states named: Alabama, Arkansas, Arizona, California, Colorado, Connecticut, Florida, Georgia, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Michigan, Minnesota, Missouri, Mississippi, North Carolina, New Hampshire, New Jersey, New Mexico, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, South Carolina, Tennessee, Texas, Utah, Virginia, Washington, Wisconsin)

How to tell if yours is affected

Model/UDI-DI: MMT-8600/00199150038978(10)1.1.0, MMT-8601/00199150038985(10)1.1.0 Software Versions: 1.0.0, 1.1.0

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2778-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedJune 16, 2026
ReportedJuly 29, 2026
CompanyMedtronic MiniMed, Inc.
Company locationNorthridge, CA, United States
Initiated byVoluntary: Firm initiated
Quantity3,465

More recalls from Medtronic MiniMed, Inc. · all medical device recalls · recalls from June 2026

Official FDA record