Recall Ledger

EOPA Elongated One-Piece Arterial Cannula BLUNT 22FR recall (Medtronic Perfusion Systems)

Medical deviceClass IIOngoing

Summary: Medtronic Perfusion Systems is recalling EOPA Elongated One-Piece Arterial Cannula BLUNT 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 77422. Reason: Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.

What is being recalled

EOPA Elongated One-Piece Arterial Cannula BLUNT 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 77422

Why

Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide - US Nationwide distribution.

How to tell if yours is affected

UDI-DI: 20763000135632, Lot Numbers: 0232446429 0232446444 0232486752 0232601515 0232601516 0232601522 0232601586 0232601591 0232601593 0232602180 0232798265 0232798353 0232849583 0232894981 0232895082 0232902694 0232902697 0232902698 0232902700 0232902803 0232902805 0232902846 0232985469 0233009922 0233108251 0233116991 0233127352 0233334923 0233334966 0233426694 0233426695 0233528727; UDI-DI: 00763000135638, Lot Numbers: 0232446444 0232798353 0232895082 0232902697 0232902803 0232902846 0232985469 0233127352 0233334923

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2806-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedJune 24, 2026
ReportedAugust 5, 2026
CompanyMedtronic Perfusion Systems
Company locationBrooklyn Park, MN, United States
Initiated byVoluntary: Firm initiated
Quantity21,290 units