EOPA Elongated One-Piece Arterial Cannula DIL TIP 22FR recall (Medtronic Perfusion Systems)
Medical deviceClass IIOngoing
Summary: Medtronic Perfusion Systems is recalling EOPA Elongated One-Piece Arterial Cannula DIL TIP 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 77622. Reason: Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.
What is being recalled
EOPA Elongated One-Piece Arterial Cannula DIL TIP 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 77622
Why
Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
Worldwide - US Nationwide distribution.
How to tell if yours is affected
UDI-DI: 20763000135670, Lot Numbers: 0232601501 0232601510 0232601549 0232601550 0232602178 0232602290 0232602302 0232910952 0232910953 0232910954 0232910955 0233244075 0233244301 0233244531 0233286362 0233286374 0233286390 0233409034 0233416071 0233528731
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-2808-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | June 24, 2026 |
| Reported | August 5, 2026 |
| Company | Medtronic Perfusion Systems |
| Company location | Brooklyn Park, MN, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 13,510 units |
More recalls from Medtronic Perfusion Systems · all medical device recalls · recalls from June 2026