Recall Ledger

EOPA 3D Arterial Cannula DIL TIP 22FR recall (Medtronic Perfusion Systems)

Medical deviceClass IIOngoing

Summary: Medtronic Perfusion Systems is recalling EOPA 3D Arterial Cannula DIL TIP 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 78222. Reason: Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.

What is being recalled

EOPA 3D Arterial Cannula DIL TIP 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 78222

Why

Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide - US Nationwide distribution.

How to tell if yours is affected

UDI-DI: 20763000135700, Lot Numbers: 0232602300 0232602383 0232612916 0232680078 0232708967 0232709001 0232910880 0232910881 0232910882 0233077983 0233245850 0233245860 0233245980; UDI-DI: 00763000135706, Lot Numbers: 0232612916 0232709001

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2810-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedJune 24, 2026
ReportedAugust 5, 2026
CompanyMedtronic Perfusion Systems
Company locationBrooklyn Park, MN, United States
Initiated byVoluntary: Firm initiated
Quantity3,621 units

More recalls from Medtronic Perfusion Systems · all medical device recalls · recalls from June 2026