REMISOL Advance with Software Version 2.5 UDI-DI code: 15099590802073 The Normand... recall (Beckman Coulter Biomedical GmbH)
Medical deviceClass IIOngoing
Summary: Beckman Coulter Biomedical GmbH is recalling REMISOL Advance with Software Version 2.5 UDI-DI code: 15099590802073 The Normand Info REMISOL Advance Data Manager collects and manages data and workflow for 128 connections. Reason: Due to software issues where does not handle combo assay results. The software may replace the final combo test result with one of the sub-results resulting in incorrect reporting of test outcomes.
What is being recalled
REMISOL Advance with Software Version 2.5 UDI-DI code: 15099590802073 The Normand Info REMISOL Advance Data Manager collects and manages data and workflow for 128 connections
Why
Due to software issues where does not handle combo assay results. The software may replace the final combo test result with one of the sub-results resulting in incorrect reporting of test outcomes.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
U.S.: FL, GA, IL, MI, and PA O.U.S.: Belgium, France, Hungary, Poland, Taiwan (Province of China), Tunisia, United Arab Emirates, and United Kingdom of Great Britain and Northern Ireland. (states named: Florida, Georgia, Illinois, Michigan, Pennsylvania)
How to tell if yours is affected
Software Version 2.5 UDI-DI code: 15099590802073
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-2834-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | April 29, 2026 |
| Reported | August 12, 2026 |
| Company | Beckman Coulter Biomedical GmbH |
| Company location | Munchen, Germany |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 20 systems |