Atellica CH Urine Albumin REF: 11537225 recall (Siemens Healthcare Diagnostics, Inc.)

Medical deviceClass IIOngoing

Summary: Siemens Healthcare Diagnostics, Inc. is recalling Atellica CH Urine Albumin REF: 11537225, in vitro diagnostic. Reason: Processing of urine albumin assay on the CI analyzers may lead to the following issues: Falsely depressed Atellica CH Urea Nitrogen (UN_c) results; Falsely depressed Atellica CH Enzymatic Hemoglobin A1c (A1c_E) results; Increased rate of reaction cuvette QC failures and Delayed results due to QC failures.

What is being recalled

Atellica CH Urine Albumin REF: 11537225, in vitro diagnostic

Why

Processing of urine albumin assay on the CI analyzers may lead to the following issues: Falsely depressed Atellica CH Urea Nitrogen (UN_c) results; Falsely depressed Atellica CH Enzymatic Hemoglobin A1c (A1c_E) results; Increased rate of reaction cuvette QC failures and Delayed results due to QC failures.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide distribution - US Nationwide including Puerto Rico and the countries of Australia, Austria, Bahamas, Belgium, Brazil, Canada, Chile, Curacao, St. Eus, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Ireland, Israel,¿Italy,¿Latvia, Malaysia, Netherlands, New Zealand, Norway, Portugal, Republic of Korea, Romania, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, United¿Kingdom.¿ (states named: Puerto Rico)

How to tell if yours is affected

All Lots within expiry/UDI: 00630414611099

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2890-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedJuly 6, 2026
ReportedAugust 12, 2026
CompanySiemens Healthcare Diagnostics, Inc.
Company locationTarrytown, NY, United States
Initiated byVoluntary: Firm initiated
Quantity7018 units