Recall Ledger

Paragonix Technologies Inc. KidneyVault Portable Renal Perfusion System REF#... recall (Paragonix Technologies Inc.)

Medical deviceClass IIOngoing

Summary: Paragonix Technologies Inc. is recalling Paragonix Technologies Inc. KidneyVault Portable Renal Perfusion System REF# PRGNX-8008-001. Reason: Due to a firmware defect, the portable organ transfer device datalogger may lose its configuration, resulting in a blank screen and loss of displayed preservation parameters.

What is being recalled

Paragonix Technologies Inc. KidneyVault Portable Renal Perfusion System REF# PRGNX-8008-001

Why

Due to a firmware defect, the portable organ transfer device datalogger may lose its configuration, resulting in a blank screen and loss of displayed preservation parameters.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

US Nationwide distribution in the states and territories of AL, AZ, CA, TN, FL, HI, MN, NJ, TX, UT, CO, CT, DC, MD, GA, KY, LA, MA, MI, MO, MS, NC, ND, NE, NM, NV, NY, PA, PR, SC, UT, VA, WI, WV, IN, OH, KY, IL, MD, MN, OR. (states named: Alabama, Arizona, California, Colorado, Connecticut, District of Columbia, Florida, Georgia, Hawaii, Illinois, Indiana, Kentucky, Louisiana, Massachusetts, Maryland, Michigan, Minnesota, Missouri, Mississippi, North Carolina, North Dakota, Nebraska, New Jersey, New Mexico, Nevada, New York, Ohio, Oregon, Pennsylvania, Puerto Rico, South Carolina, Tennessee, Texas, Utah, Virginia, Wisconsin, West Virginia)

How to tell if yours is affected

All Serial Numbers/UDI: 00850005470342 Software Version: Paragonix App Ver. 6.1.1/6.1.2

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2967-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedJuly 20, 2026
ReportedAugust 26, 2026
CompanyParagonix Technologies Inc.
Company locationBraintree, MA, United States
Initiated byVoluntary: Firm initiated
Quantity1445 systems

More recalls from Paragonix Technologies Inc. · all medical device recalls · recalls from July 2026

Official FDA record