MiniMed 640G Insulin Pump recall (Medtronic MiniMed, Inc.)

Medical deviceClass IIOngoing

Summary: Medtronic MiniMed, Inc. is recalling MiniMed 640G Insulin Pump, Model: MMT-1711, MMT-1712;. Reason: Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis.

What is being recalled

MiniMed 640G Insulin Pump, Model: MMT-1711, MMT-1712;

Why

Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

International distribution to the countries of Germany, Denmark, Austria, Netherlands, United Kingdom, Italy, Belgium, Ireland, Switzerland, Sweden, Northern Ireland.

How to tell if yours is affected

Model/UDI-DI/Serial Numbers: MMT-1711/00763000404550/NG2660048H, NG2663343H, NG3015306H; MMT-1712/00763000404567/NG2628486H, NG2758213H, NG2999651H, NG3349198H, NG3351719H, NG2678842H, NG2831821H, NG2831546H, NG2932027H, NG3049657H, NG3033094H, NG2831538H, NG3738739H

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2980-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedJuly 13, 2026
ReportedAugust 26, 2026
CompanyMedtronic MiniMed, Inc.
Company locationNorthridge, CA, United States
Initiated byVoluntary: Firm initiated
Quantity16

More recalls from Medtronic MiniMed, Inc. · all medical device recalls · recalls from July 2026