Medline breathing circuits labeled as: 1) MEDLINE HUDSON RCI recall (Medline Industries, LP)

Medical deviceClass IIOngoing

Summary: Medline Industries, LP is recalling Medline breathing circuits labeled as: 1) MEDLINE HUDSON RCI, CIRCUIT, NEO, SING LIMB, HTD,COLUMN, Medline Product Number/SKU HUD78001KIT (Teleflex Product Number/SKU 780-01KIT); 2) MEDLINE HUDSON RCI, CIRCUIT, NEO, DUAL LIMB HTD, COLUMN, Medline Product Number/SKU HUD78007KIT (Teleflex Product. Reason: Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.

What is being recalled

Medline breathing circuits labeled as: 1) MEDLINE HUDSON RCI, CIRCUIT, NEO, SING LIMB, HTD,COLUMN, Medline Product Number/SKU HUD78001KIT (Teleflex Product Number/SKU 780-01KIT); 2) MEDLINE HUDSON RCI, CIRCUIT, NEO, DUAL LIMB HTD, COLUMN, Medline Product Number/SKU HUD78007KIT (Teleflex Product Number/SKU 780-07KIT); 3) MEDLINE HUDSON RCI, DBD-CIRCUIT, NEO, DUAL LIMB HTD, COLUMN, Medline Product Number/SKU HUD78008KIT (Teleflex Product Number/SKU 780-08KIT); 4) MEDLINE HUDSON RCI, DBD-CIRCUIT, NEO, DUAL LIMB HTD, COLUMN, Medline Product Number/SKU HUD78009KIT (Teleflex Product Number/SKU 790-09KIT); 5) MEDLINE HUDSON RCI, CIRCUIT, NEO, DUAL LIMB, HTD,COLUMN, Medline Product Number/SKU HUD78017KIT (Teleflex Product Number/SKU 780-17KIT); 6) MEDLINE HUDSON RCI, CIRCUIT, NEO, DUAL HTD, DUAL LIMB,COLUMN, Medline Product Number/SKU HUD78018KIT (Teleflex Product Number/SKU 780-18KIT); 7) MEDLINE HUDSON RCI, CIRCUIT,ISO-SHIELD SINGLE LIMB,INFANT, Medline Product Number/SKU HUD88001KIT (Teleflex Product Number/SKU 880-01KIT); 8) MEDLINE HUDSON RCI, CIRCUIT,ISO-SHIELD,INFANT,HTD,COLUMN,DRN, Medline Product Number/SKU HUD88013KIT (Teleflex Product Number/SKU 880-13KIT); 9) MEDLINE HUDSON RCI, ISO-SHIELD CIRCUIT,INFNT,HTD,DL,DRAIN, Medline Product Number/SKU HUD88014KIT (Teleflex Product Number/SKU 880-14KIT); 10) MEDLINE HUDSON RCI,ISO-SHIELD CIRCUIT,INFNT,HTD,DL,DRAIN, Medline Product Number/SKU HUD88015KIT (Teleflex Product Number/SKU 880-15KIT); 11) MEDLINE HUDSON RCI, ISO-SHIELD CIRCUIT,INFNT,HTD,DL,DRAIN, Medline Product Number/SKU HUD88017KIT (Teleflex Product Number/SKU 880-17KIT).

Why

Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide distribution - US Nationwide and the country of Panama.

How to tell if yours is affected

1) Medline Product Number/SKU HUD78001KIT (Teleflex Product Number/SKU 780-01KIT), UDI/DI 10197344010752 (each) 20197344010759 (case), All lot numbers; 2) Medline Product Number/SKU HUD78007KIT (Teleflex Product Number/SKU 780-07KIT), UDI/DI 10197344010738 (each) 20197344010735 (case), All lot numbers; 3) Medline Product Number/SKU HUD78008KIT (Teleflex Product Number/SKU 780-08KIT), UDI/DI 10197344010721 (each) 20197344010728 (case), All lot numbers; 4) Medline Product Number/SKU HUD78009KIT (Teleflex Product Number/SKU 790-09KIT), UDI/DI 10197344010714 (each) 20197344010711 (case), All lot numbers; 5) Medline Product Number/SKU HUD78017KIT (Teleflex Product Number/SKU 780-17KIT), UDI/DI 10197344010677 (each) 20197344010674 (case), All lot numbers; 6) Medline Product Number/SKU HUD78018KIT (Teleflex Product Number/SKU 780-18KIT), UDI/DI 10197344010905 (each) 20197344010902 (case), All lot numbers; 7) Medline Product Number/SKU HUD88001KIT (Teleflex Product Number/SKU 880-01KIT), UDI/DI 10197344010929 (each) 20197344010926 (case), All lot numbers; 8) Medline Product Number/SKU HUD88013KIT (Teleflex Product Number/SKU 880-13KIT), UDI/DI 10197344010493 (each) 20197344010490 (case), All lot numbers; 9) Medline Product Number/SKU HUD88014KIT (Teleflex Product Number/SKU 880-14KIT), UDI/DI 10197344010486 (each) 20197344010483 (case), All lot numbers; 10) Medline Product Number/SKU HUD88015KIT (Teleflex Product Number/SKU 880-15KIT), UDI/DI 10197344010479 (each) 20197344010476 (case), All lot numbers; 11) Medline Product Number/SKU HUD88017KIT (Teleflex Product Number/SKU 880-17KIT), UDI/DI 10197344010462 (each) 20197344010469 (case), All lot numbers;

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2993-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedJuly 13, 2026
ReportedSeptember 2, 2026
CompanyMedline Industries, LP
Company locationNorthfield, IL, United States
Initiated byVoluntary: Firm initiated
Quantity271406 units