Alphenix Models INFX-8000C INFX-8000F INFX-8000V This device is a digital... recall (Canon Medical System, USA, INC.)
Medical deviceClass IIOngoing
Summary: Canon Medical System, USA, INC. is recalling Alphenix Models INFX-8000C INFX-8000F INFX-8000V This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen . Reason: Due to the fixing screws of ceiling movement gear in ceiling suspended arm that are part of the X-ray interventional systems may become loose.
What is being recalled
Alphenix Models INFX-8000C INFX-8000F INFX-8000V This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.
Why
Due to the fixing screws of ceiling movement gear in ceiling suspended arm that are part of the X-ray interventional systems may become loose.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
U.S. Nationwide distribution in the states of CA, GA, MN, NC, NV, PA, OH, and TX. (states named: California, Georgia, Minnesota, North Carolina, Nevada, Ohio, Pennsylvania, Texas)
How to tell if yours is affected
Model Number: INFX-8000C/B3 UDI code: Pending SID Number:30101165 Serial Number: B3E2422005 Model Number: INFX-8000F/SG UDI code: Pending SID Number: 30093711 Serial Number: SGC2412001 Model Number: INFX-8000V/SZ UDI code: Pending SID Number: 30029686 Serial Number: SZA2422001 SID Number: 79000002 Serial Number:SZA2442004 Model Number: INFX-8000V/SX UDI code: Pending SID Number: 30049491 Serial Number: SXA2432001 SID Number: 30100053 Serial Number: SXA2432002 Model Number: INFX-8000V/B9 UDI code: Pending SID Number: 30059175 Serial Number: B9F2412001 Model Number: INFX-8000V/ST UDI code: Pending SID Number: 30059180 Serial Number: STA2412001 Model Number: INFX-8000V/SZ UDI code: Pending SID Number: 30095017 Serial Number: SZA2432003 SID Number: 30097052 Serial Number: SZA2422002
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-3014-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | August 4, 2026 |
| Reported | September 2, 2026 |
| Company | Canon Medical System, USA, INC. |
| Company location | Tustin, CA, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 10 systems |