Recall Ledger

Product Name: DermaV Laser System Model/Catalog Number: DermaV Medical laser... recall (Lutronic Corporation)

Medical deviceClass IIOngoing

Summary: Lutronic Corporation is recalling Product Name: DermaV Laser System Model/Catalog Number: DermaV Medical laser device that delivers pulsed laser energy at wavelengths of 532 nm and 1064 nm through a fiber-optic handpiece. The system may be configured with an Intelligent Cooling Device (ICD) to provide epidermal cooling during la. Reason: Utilizing higher fluences at 1064nm wavelength with current cryogen parameters may result in a spark/loud noise during treatment, potentially resulting in patient burn.

What is being recalled

Product Name: DermaV Laser System Model/Catalog Number: DermaV Medical laser device that delivers pulsed laser energy at wavelengths of 532 nm and 1064 nm through a fiber-optic handpiece. The system may be configured with an Intelligent Cooling Device (ICD) to provide epidermal cooling during laser treatment

Why

Utilizing higher fluences at 1064nm wavelength with current cryogen parameters may result in a spark/loud noise during treatment, potentially resulting in patient burn.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide distribution - US Nationwide and the countries of Canada and European Union.

How to tell if yours is affected

Lot Code: Model: DermaV Software Version: 1.06 UDI #: 08800291820200 Serial Numbers: DV025A010, DV025A011, DV025A012, DV025A013, DV025A014, DV025A015 DV025A016, DV025A018, DV025A019, DV025A020, DV025A021, DV025A022, DV025B011, DV025B012, DV025B013, DV025B014, DV025B015, DV025D006, DV025D007, DV025D008, DV025D009, DV025D010, DV025E003, DV025E004, DV025E005, DV025E006, DV025E007, DV025E008, DV025E009, DV025E010, DV025E011, DV025E012, DV025E013, DV025E014, DV025E016, DV025E017, DV025F012, DV025F013, DV025F014, DV025F015, DV025F016, DV025G011, DV025G012, DV025G013, DV025G014, DV025G015, DV025G016, DV025G017, DV025G018, DV025G019, DV025G020, DV025J021, DV025J022, DV025J023, DV025J025, DV025J026, DV025J027, DV025J028, DV025J034, DV025J035, DV025J036, DV025J037, DV025J038, DV025J040, DV025K018, DV025K019, DV025K020, DV025K021, DV025K023, DV025K024, DV025K025, DV025K027, DV025M031, DV025M032, DV025M033, DV025M034, DV026B001, DV026B002, DV026B003, DV026B004, DV026B005, DV026B010, DV026B014, DV026B015, DV026B016, DV026B017, DV026B018, DV026C006, DV026C007, DV026C008, DV026C009, DV026C010, DV025M035

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-3110-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedAugust 6, 2026
ReportedSeptember 16, 2026
CompanyLutronic Corporation
Company locationGoyang, Korea (the Republic of)
Initiated byVoluntary: Firm initiated
Quantity1065

More recalls from Lutronic Corporation · all medical device recalls · recalls from August 2026