Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04 recall (GE Medical Systems Information Technologies Inc)
Medical deviceClass IIOngoing
Summary: GE Medical Systems Information Technologies Inc is recalling Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04, 5867474-13 Product code STU CSCS V3 MAI800 Integrated Part Number 5867474-03 Product Code STV. Reason: Device may not resume patient monitoring for at least 3 minutes following system restart. Issue can lead to potential delay in recognition of clinically significant events.
What is being recalled
Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04, 5867474-13 Product code STU CSCS V3 MAI800 Integrated Part Number 5867474-03 Product Code STV
Why
Device may not resume patient monitoring for at least 3 minutes following system restart. Issue can lead to potential delay in recognition of clinically significant events.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
Worldwide - US Nationwide distribution.
How to tell if yours is affected
5867474-04 UDI-DI 00195278343512 Device Serial Numbers begin with STU 5867474-13 UDI-DI 00195278516367 Device Serial Numbers begin with STU 5867474-03 UDI-DI 00195278343383 Device Serial Numbers begin with STV
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-3143-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | July 30, 2026 |
| Reported | September 23, 2026 |
| Company | GE Medical Systems Information Technologies Inc |
| Company location | Waukesha, WI, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 29,570 (23,632 US; 5,938 OUS) |