Recall Ledger

Philips EPIQ Elite Diagnostic Ultrasound System Model # 795234 recall (Philips Ultrasound, LLC)

Medical deviceClass IIOngoing

Summary: Philips Ultrasound, LLC is recalling Philips EPIQ Elite Diagnostic Ultrasound System Model # 795234. Reason: Due to software issue with ultrasound system, doppler sample is displayed shallower than its actual measurement location, potentially causing inaccurate blood flow velocity measurements and misdiagnosis.

What is being recalled

Philips EPIQ Elite Diagnostic Ultrasound System Model # 795234

Why

Due to software issue with ultrasound system, doppler sample is displayed shallower than its actual measurement location, potentially causing inaccurate blood flow velocity measurements and misdiagnosis.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide - US Nationwide distribution in the states of AL, CA, CT, FL, GA, KS, MD, MN, NJ, NY, OH, PA, SC, TN, TX, VA, WA, WI and the countries of Austria, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lithuania, Netherlands, Norway, Philippines, Poland, Romania, Slovenia, South Korea, Spain, Sweden, Switzerland, Taiwan , United Kingdom, Vietnam. (states named: Alabama, California, Connecticut, Florida, Georgia, Kansas, Maryland, Minnesota, New Jersey, New York, Ohio, Pennsylvania, South Carolina, Tennessee, Texas, Virginia, Washington, Wisconsin)

How to tell if yours is affected

All Philips EPIQ Elite Ultrasound Systems running software version 14.0, UDI(s): (01)00884838102514 (01)00884838107540 (01)00884838114159 (01)00884838124912 (01)00884838128125

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-3249-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedAugust 7, 2026
ReportedSeptember 30, 2026
CompanyPhilips Ultrasound, LLC
Company locationBothell, WA, United States
Initiated byVoluntary: Firm initiated
Quantity219 units

More recalls from Philips Ultrasound, LLC · all medical device recalls · recalls from August 2026