Computed Tomography X-ray System recall (Shanghai United Imaging Healthcare Co., Ltd.)

Medical deviceClass IIOngoing

Summary: Shanghai United Imaging Healthcare Co., Ltd. is recalling Computed Tomography X-ray System, Model: uCT ATLAS with software. Reason: Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.

What is being recalled

Computed Tomography X-ray System, Model: uCT ATLAS with software

Why

Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

US Nationwide distribution in the states of CA, PR, UT, WV, ND, NE, NC, TX, ID, LA, MT, FL, CO, IN, KY, AL, OH, MI, IA, MO, NY, NV, TN, WV, SC, WI, IL, MS, MA, NJ. (states named: Alabama, California, Colorado, Florida, Iowa, Idaho, Illinois, Indiana, Kentucky, Louisiana, Massachusetts, Michigan, Missouri, Mississippi, Montana, North Carolina, North Dakota, Nebraska, New Jersey, Nevada, New York, Ohio, Puerto Rico, South Carolina, Tennessee, Texas, Utah, Wisconsin, West Virginia)

How to tell if yours is affected

UDI-DI: 06971576831104. Serial Number/Software: 863001/R001.17.6.1607336-Re3100SP4, 863002/R001.17.6.1607336-Re3100SP4, 863004/R001.17.6.1607336-Re3100SP4, 863005/R001.17.6.1607336-Re3100SP4, 863007/R001.21.0.1897081-Re3310SP1, 863008/R001.21.0.1897081-Re3310SP1, 869001/R001.22.0.2012850-Re3320, 869002/R001.19.4.1810621-Re3300SP1H1_P3, 869003/R001.21.0.1897081-Re3310SP1, 869004/R001.21.0.1897081-Re3310SP1, 11CT863010/R001.17.6.1607336-Re3100SP4

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-3281-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedAugust 11, 2026
ReportedOctober 7, 2026
CompanyShanghai United Imaging Healthcare Co., Ltd.
Company locationShanghai, China
Initiated byVoluntary: Firm initiated
Quantity11

More recalls from Shanghai United Imaging Healthcare Co., Ltd. · all medical device recalls · recalls from August 2026

Official FDA record